Code No: MS-5050-40-11-M
Technical Specifications
*Reduces gas from a maximum inlet pressure of 230 bar to levels of 10 bar, 20 bar, or 40 bar.
*Compatible with cylinders or cylinder pallets via two valve-integrated cylinder connections.
*Additional cylinders or cylinder pallets can be connected to the system through extension ports using optional valve-integrated connectors.
*All system components are mounted on high-strength stainless steel sheets.
*In oxygen systems, a flame arrestor is installed before the ball valve on the outlet line.
*A manual purge valve located after the regulator allows for safe evacuation of gas before cylinder replacement or maintenance operations.
*Check valves located behind each cylinder valve prevent backflow, enhancing system safety.
*A pressure gauge positioned just before the outlet allows monitoring of line pressure, even when the main valve is closed.
*From the regulator outlet to the system's end point, the line is piped with Ø28 mm copper tubing.
Safety relief valves installed after the pressure regulator ensure safe discharge of excess pressure in case of malfunction.
.gif)