Semi-Automatic Oxygen Manifold Systems

Semi-Automatic Oxygen Manifold Systems
Code No: MS-50501-10-11-YO-AE
Semi-Automatic Oxygen Manifold Systems
Gas Type :Oxygen
Inlet Gas Pressure :230 Bar
Outlet Gas Pressure :10 Bar
Inlet Connection :W21.8x1/14"
Outlet Connection :G3/4"
Technical Specifications
*A maximum inlet pressure of 230 bar is reduced to outlet levels of 10 bar, 20 bar, or 40 bar.
*Equipped with 2 cylinder connections with valves, suitable for use with individual cylinders or cylinder bundles.
*Extension connections on both inlet sides allow additional cylinders or bundles to be connected to the system by purchasing extra cylinder valve connections.
*While one side of the system is operating, the other side remains on standby. When the cylinder or bundle is emptied, the gas flow automatically switches to the other side, ensuring uninterrupted operation.
*The central lever can be turned upward or downward to select the active gas line.
*All components of the system are mounted on a high-strength stainless steel sheet.
*Thanks to the vent valves located on the inlet sides, the gas inside the system can be safely discharged before cylinder replacement or maintenance.
*In oxygen systems, a flashback arrestor is installed before the ball valve on the outlet line.
*Check valves located at both cylinder connection valves prevent gas backflow, thereby enhancing system safety.
*Thanks to its compact design, it occupies minimal space while its cost-effective structure reduces investment expenses.
*The safety relief devices located in the pressure regulators ensure safe venting of excess pressure in case of malfunction.